Test voltage: 60-600V: programmable in 0.1 Volt steps

Test current: 2.5-50A full load programmable in 0.001A steps light load 10% of full load setting

Power factor test angle: 0-359.9 degrees selectable in 0.1 degree increments

Test revolutions:
• Series full load selectable 1-65,534*
• Elements full load 1-65,534
• Series light load selectable 1-65,534*
• Elements light load 1-65,534
• Series power factor selectable 1-65,534*
• Elements power factor 1-65,534

Demand (KW) testing: Standard revolutions (1/999 to 99999 revolutions) time run (up to 99 hrs, 59 min, 59 sec)

Contact device testing: Form "A" or "C" capability

Solid-state voltage and current loading

Voltage and current electronically ramped between tests

Automatic slew between tests to minimize test time

Potential clips closed for testing

Smart Socket™ with minimal insertion force

Electronic power amplifier safety shutdown circuitry

Modulated laser optics for through hole and reflect disk sensing

No sensitivity adjustment necessary for optics

Three optical pickups

Visual and audible (volume controlled) pulse indication for aid in aligning optical sensor

Radian Research™ RD-20 standard - easily removed for certification

Provided PC:
2.6 GHz Processor, 3.5" Floppy Disk Drive • 40 GB HD, 128 MB RAM, 2 USB Ports, Printer Port, CD ROM, • 17" SVGA Monitor - .26 dot pitch (1600 X 1200 - 256 colors) • Software Package: Windows™ XP Professional, 101 Key Keyboard, Mouse

Winboard™ software, Windows™ 98, 98SE, ME, 4.0 NT, 2000 NT, and XP compatible

WATT-Net data acquisition system

One (1) standard per meter position, optionally expandable to three (3)

Front panel controls allow fast, easy test selection


*Selectable Per Element